MGA Prime
MGA Prime
  • Product Code: MGA Prime
  • Availability: In Stock
  • $15,139.00

  • Ex Tax: $15,139.00
Request Discount Price

For long time measurement of industrial combustion, large boilers, gas engines and turbines, furnaces and many more

  • Precise and very stable offset measuring technique using special non dispersive infrared (NDIR) technique, for enhanced long time measurements
  • Up to 8 gas NDIR measurement:
    CO/CO2/C3H8/CH4/SO2/NO/NO2/N2O and oxygen with either long-life O2-ECS or
    O2-PM (paramagnetic cell)
  • Remote display through MRU4U App over bluetooth on smartphone or tablet or trough VCN server via LAN/WLAN
  • 7” high-resolution (800×480 pixel) TFT color display with swipe and touch technique as for smartphones, with LINUX operating system
  • Intuitive icons and country specific language on display
  • Modern and robust design, Aluminum enclosure with rubber molded impact
    protection for use in harsh industrial environment
  • Advanced algorithms to control analyzer’s operation, like warming up, compensation of cross-sensitivity values, response time – all which have impact on the quality and performance
  • Measurement of many additional parameters:
    temperatures (ambient and flue gas), pressures (atmospheric and differential),
    internal sample flow monitoring
  • Measurement of flue gas flow velocity (using of Pitot tube) and calculation of
    stack gas flow rate
  • Calculation of all combustion parameters like:
    heat loss, combustion efficiency, excess air coefficient, dew point temperature and others
  • Results stored in database – possibility to view entire measurement session in graphical form, create measurement reports, export data to CSV files, and more
  • PC communication via LAN – Ethernet
  • Wireless data transfer via Bluetooth dongle or WLAN dongle
  • Long distance data transfer via RS485 cable
  • Analog outputs (8-channels), analog inputs (4-channels)
  • SAMPLE GAS CONDITIONING: Dual gas cooler (Peltier) with dual condensate
    draining pumps, with constant dew point +5°C
    when connected to mains power supply
  • Condensate monitoring and alarm on display
  • Strong gas sampling pump, regulated flow of approx 1l/min
  • Gas sampling probe heated, with exchangeable probe tubes, including K-type thermo-couple for flue gas temperature measurement
  • Heated sampling line, Teflon 6/4mm, 9.84′ or 16.4′ or 32.81′ (3 m – 5 m – 10 m) and other lengths
TECHNICAL DATA
Gas measurementMethodMeas. range

(0…min / max)

ResolutionAccuracy **
O2 – OxygenECS0 to 25.00 %0.01 %0.2 %
O2 – OxygenPM0 to 25.00 %0.01 %0.1 %
CO2 – Carbon monoxideNDIR0 to 5 / 40%0.01 %± 0.2 % or 2 % reading
CO – Carbon dioxideNDIR0 to 200 / 10,000 ppm1 ppm± 4 ppm or 2 % reading
HC – Hydrocarbons as CH4NDIR0 to 500 / 10,000 ppm1 ppm± 10 ppm or 2 % reading
HC – Hydrocarbons as C3H8NDIR0 to 200 / 10,000 ppm1 ppm± 10 ppm or 2 % reading
NO – Nitric oxideNDIR0 to 250 / 4,000 ppm1 ppm± 5 ppm or 2 % reading
NO2 – Nitrogen dioxideNDIR0 to 200 / 1,000 ppm1 ppm± 4 ppm or 2 % reading
SO2 – Sulfur dioxideNDIR0 to 200 / 4,000 ppm1 ppm± 4 ppm or 2 % reading
N2O – Nitrous oxideNDIR0 to 200 / 1,000 ppm1 ppm± 4 ppm or 2 % reading
OTHER MEASUREMENTS / CALCULATIONSMethodMeasuring rangeResolutionAccuracy **
Tgas – stack gas temperatureNiCrNi32°F to 2.192°F K-type1°F± 2°F or 2 % reading
Tair – combustion air temperatureNiCrNi32°F to 932°F K-type1°F± 2°F or 2 % reading
Tamb – ambient air temperatureRTD32°F to 212°F PT20001°F± 2°F or 2 % reading
Diff.pres. – differential pressurePiezoresistive-48 to +48 inH2O1 Pa± 2 Pa or 1 % reading
GasFlow – flow velocity – differential pressurePitot3 … 100 m/s1 m/s± 1 m/s or 1 % reading
AUX – auxiliary connector –

standardized external signal

Softwarefor K-thermocouple, 0 … 10 Vdc , 4 … 20 mA, RS485
Combustion calculations – fuel type dependentSoftwareheat loss, ExcAir, Air Ratio (λ), dewpoint
Emission calculationsSoftwaremg/Nm3 , reference to O2 , g/kWh, g/s , kg/h
GENERAL TECHNICAL DATA 
Operating systemLINUX
Display, operation7“ TFT (800 x 480 px) colour display, backlit, with touch pad
Data storage type /capacityinternal 10.000 data sets, external USB stick
Interface to external PC/notebookEthernet, Bluetooth, WLAN, RS485
Cable communication interfaceRS485, RJ45 (Ethernet)
Wireless communicationBluetooth, WLAN
Thermal printerexternal only
Analog output 4 – 20 mA/analog input 4 – 20 mA8 channel out / 4 channel in/user configurable
Universal analog input – AUX –0…10 Vdc / 4…20 mA / NiCrNi / RS485
System warming up time30 minutes (typical)
Warming up temperature NDIR bench131°F (55 °C)
Mains free operation time / stand-by only1 hour
Internal batteryLi-Ion , 96W, for standby
Operating conditions41°F to 113°F, RH up to 95% non condensing
Storage temperature-4°F to 122°F
Power supply / consumption86 to 265 Vac / 47…63 Hz / 105 W (analyzer only)
Enclosure materialaluminum, rubber molded impact protection
Protection classIP20 (or IP42 inside transport case)
Dimensions16.92“ x 11.41“ x 5.9“ (WxHxD)
Weightfrom 16.5 lbs. for minimal configuration

* overload range of ECS is usable only for short duration ** which ever is larger

 

Reviews (0)

Write a review

Note: HTML is not translated!
         
RELATED PRODUCTS

  • MGA 5

    MGA 5

    Professional and portable analyzer for high quality combustion analysis during process gas monitoring. Ideal for use when maintain..

  • MGA Luxx

    MGA Luxx

    Ideal for long time measurement of industrial combustion, large boilers, gas engines and turbines, furnaces and man..